11.4 Brief operating instructions

1. Switch on the system at the mains power switch down on the right side

of the unit.

2. Switch on the cooling trap and/or the vacuum pump.
3. Close the lid.
4. Starting the warm-up phase:

• Enter the set values for the warm-up phase into "menu" – "options" –

"Timer warm-up" und "Temperature warm-up", confirm your entries
and quit the menu.

• Press the left-hand function key "mode".
• Select the "Warm-up" menu and confirm.
• Open the lid when the warm up phase has ended.
• Install the rotors and close the lid.

5. Changing the set values for time, temperature, vacuum and safety

pressure:
• Select "menu" with the right-hand function key.
• Select "Change set values for manual mode" from the main menu.
• Select the set value by using the up/down keys and confirm with the

right-hand function key "edit".

• Change the set value by using the up/down keys and confirm with

the right-hand function key "ok".

• Quit the menu with the left-hand function key.

6. Starting the process "Evaporation manual" or "Evaporation program":

• Open the menu "Start with phase..." by pressing the left-hand

function key "mode".

• Select "Evaporation manual" or "Evaporation program" using the

up/down keys.

• Start the evaporation process by pressing the right-hand function

key.

• The function "Evaporation program" can only be selected if a

program has been created before.

• The start of the evaporation process starts the following automatic

process run:
- The rotor speed increases (rotating direction left),
- At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid delays in boiling),
- At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump,
- Further speed increase until the preselected maximum rotor speed
is reached.

7. Interrupting an evaporation process:

• Press the left-hand function key "mode".
• Select "Stop evaporation" and confirm.

Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:

• At maximum speed: the stop valve closes the connection to the vacuum

pump, the aeration valve opens; the rotor speed is maintained for 30
sec (aeration phase, to avoid delays in boiling).