8 Maintenance and service

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

59

Translation of the original operating manual

Pos: 177 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 33 IR /08 0 War tung un d Inst and haltu ng/ 080 -00 10- 0020 Rot ork amm er RVC 2-3 3 IR @ 31\ mod _14 074 840 934 90_6 8.d ocx @ 209 100 @ 3 @ 1

8.1.2 Rotor chamber

If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion.

Cleaning of the rotor chamber
• Open the lid.
• Disconnect the power cord from the wall outlet or instrument receptacle.

DANGER

Hot surface!
Let the rotor chamber cool down to room temperature!


1

Rotor chamber

2

Electronic
temperature
measurement system

of the shaft

3

Cap

4

Rotor shaft

5

Magnetic coupling on

rotor side

Fig. 52: Rotor chamber with rotor shaft incl. magnet coupling

• To get access to the bottom of the unit, turn the unit upside down or

place the unit between two trestles or two tables with a sufficient
number of people helping you.

CAUTION

The rotational vacuum concentrator weighs approx. 49 kg!