7 Malfunctions and error correction

52

Version 08/2009, Rev. 2.1 of 02/12/2014 • sb

Translation of the original operating manual

Pos: 153 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 Stör ung en u nd Fehl ersuc he === == == === == == === == == === == = @ 25\ mod _140 498 518 589 4_6 8.doc x @ 1 874 15 @ 1 @ 1

7 Malfunctions and error correction

Pos: 154 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10 Allg emei ne Stö run gen @ 2 5\mo d_1 404 985 186 669 _68. docx @ 18 742 9 @ 2 @ 1

7.1 General malfunctions

Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.4 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function

key.

Pos: 155 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 156 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 157 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 001 0 Fehle rm eldun gen (a uße r 2 -18 ) @ 25\ mod_ 140 498 518 767 1_6 8.doc x @ 1 874 43 @ 3 @ 1

7.1.1 Error messages

Error

Possible reason

Correction

No vacuum sensor signal

• The control unit receives an

invalid value from the vacuum

sensor.

• Check the connecting cables

and plug-in connections of the
vacuum sensor. If necessary,

replace the vacuum sensor.

No rotor chamber sensor signal

• The temperature sensor of the

rotor chamber is disconnected

or defective

• Contact the after-sales service

(see chapter 7.2 - "Service

contact")

No rotor sensor signal

• The rotor temperature sensor is

disconnected or defective

• Contact the after-sales service

(see chapter 7.2 - "Service

contact")

No ice condenser sensor signal

• The data communication link to

the cooling trap is interrupted,
or the ice condenser
temperature sensor is

disconnected or defective

• Check the data link between the

rotational vacuum concentrator

and the cooling trap

Lid cannot be opened

• The rotor chamber is not

aerated completely

• Aerate the rotor chamber

completely

Lid contacts error

• The system has detected an

open lid during evaporation.
The rotor will be stopped
immediately. The lid must be

closed while the rotor is moving.

• Check the lid contacts. If

necessary, contact the after-
sales service (see chapter 7.2 -

"Service contact")

Rotor overtemperature

• The maximum permissible rotor

temperature of 85°C has been

exceeded

• If the system does not cool

down, disconnect it immediately
from the power supply. Contact
the after-sales service (see

chapter 7.2 - "Service contact")

Chamber overtemperature

• The maximum permissible

chamber temperature has been

exceeded

• If the system does not cool

down, disconnect it immediately
from the power supply. Contact
the after-sales service (see

chapter 7.2 - "Service contact")