Casella CEL-35X dBadge Users Handbook – Page 14


The subsequent screens will cycle through results from the last
measurement run, as shown in Figure 10. Values shown will be
dependent on whether the CEL-35X is configured to display ISO or
OSHA parameters (see sections 5.6 & 5.7).

ISO VIEW

DURATION

07:45:12

LAEQ 89.9

LCPK 101.4

PA Hrs

3.20

PROJ DOSE

352.5 %

OSHA VIEW

DURATION

07:59:32

LAVG 111.4

LZPK 119.4

OSHA DOSE

114.3 %

PROJ DOSE

175.3 %

Additional screen on CEL-352 dB

adge „Plus‟ model:


Figure 10

Note:

If „Pro Mode‟ is enabled additional menus will be displayed, see

section 5.5.

The over-range symbol

will be shown if the CEL-35X has been

exposed to noise over the linear operating range. The screens will
continue to automatically cycle through as shown in Figure 9 and 10
until another action is performed. Pressing the „R‟ key will stop the
screens cycling for 5 seconds. Screens can also be manually cycled
through by

pressing the „R‟ key repeatedly. Note that if the memory is

currently empty the screens shown in Figure 10 will not be shown.

5.2 Calibration

It is important to calibrate each dBadge before and after use, in
accordance with workplace noise regulations. The dBadge records
calibration levels and times which can be viewed later on Casella
insight data management software.

Note that the dBadge will not enter calibration mode if a measurement
run is taking place. If a run is in progress, stop the run according to
section 5.4.

LCEQ 92.4

dB

LC-A 4.8

dB