2 microphone specification (cel-252) – Casella CEL CEL-35X dBadge User Manual

Page 32

Advertising
background image

Casella CEL-35X dBadge Users Handbook – Page 32

Indication

Overload Indication:

140.4dB RMS (i.e. 0.1dB above the top of the
range),

Peak Overload Value: 143.4dB Peak Linear (i.e. 3.1dB above the top

of the range),

Stabilisation Time:

3 seconds after power up,


Operating Environment
Temperature:

0

o

C to +40

o

C,

Pressure:

65 - 108kPa,

Humidity ± 0.5dB:

30% - 90% (non-condensing),

Magnetic Fields:

Negligible,


Storage Environment
Temperature:

-10

o

C to +50

o

C,


Battery:

Internal NiMH, 28 hour battery life,

Charge Time:

Less than 90 minutes from discharged,

Dimensions mm (in): 72x47x52 (2.8x1.8x2.0),
Weight g (oz):

68 (2.4),


The

CEL-35X/IS

Intrinsically

Safe

dBadge

is

labelled

with the following:
ATEX:
I M1
II 1G
EEx ia IIC T2 (222

o

C)

EEx ia I
(T

a

= -20

o

C to +40

o

C)

7.2 Microphone Specification (CEL-252)


Classification:

Type 2

Nominal Open-Circuit Sensitivity (250Hz):

-28dB ± 3.0dB
(30mV/Pa)

Polarization voltage:

0V

Capacitance (polarized, 250Hz):

11-15pF

Operating temperature range:

0 to +40

o

C

Cartridge thermal noise:

25dB(A)

Dimensions:

IEC 61094-4 Type WS
2

FM/CSA:
Class 1
Division 1
Groups A, B, C and D.
Temperature classification T2

Advertising