HB3256-03

Page 15

Temperature and Humidity Sensor Wiring.

The combined temperature and humidity sensor is supplied with a radiation screen. The unit
should be secured to a suitable support structure free from possible shading effects or error
sources.

The temperature / humidity probe is connected to the APM system as detailed below.

Sensor connection

Sensor wire
colour

APM system connection

Supply +ve

RED

12 volt supply (on terminal strip)

Supply 0v

BLUE

0 volt reference (on terminal strip)

Temperature signal

YELLOW

CH1 +

Humidity signal

GREEN

CH1 -

Screen

Yellow/Green

Chassis earth point

Before applying power to the system, confirm all system wiring is correct.