Casella CEL Airborne particulate monitoring system User Manual

Page 52

Advertising
background image

HB3256-03

Page 50

Power Supply: 110 / 230 v. a.c. 50-60Hz at typically 120 Watts.

Operating temperature: -10

°C to +40°C, 0 to 95% RH, non condensing.

Enclosure Dimensions: WxHxD =536x747x300 mm.

Inlet tube: Diameter 33.8mm, Length 388mm ( 750mm max length).

Enclosure: IP65 rating with wall mount brackets or galvanised steel frame.

Wind speed (optional): a 3 cup anemometer to measure wind speed and windrun.
Construction: anodised aluminium alloy.
Operating range: <0.4m/s to 75m/s.
Distance constant: 5.3(typ).
Accuracy:

±1% above 3m/s.

Wind direction (optional): Wind direction sensor utilising a low friction wire wound
potentiometer.
Construction: anodised aluminium alloy.
Distance constant: 2.3(typ).
Electrical angle: 0-355

°.

Max. wind speed: 75m/s.
Accuracy: 2 degrees in steady winds over 5m/s.

Temp/RH (optional): a combined sensor supplied fitted with a protective radiation shield.
Construction: plastic radiation shield, ABS case to IP65.
Humidity: capacitive sensor.
Range: 0 to 1v. output for 0 -100%.
Calibrated accuracy:

±2%.

Linearity: 5 to 95% RH

±1.5%.

Temperature: PRTD thin film sensing element.
Range: 0 to 1v. output for -40 to 60

°C.

Accuracy:

±0.3°C at 0°C, ±0.5°C at 50°C.