1.4.3.1 Analog input/output signals

Analog input signals should be connected to pin 2 (+) and 0 V/common.
Analog ouput signals are available at pin 1 (+) and 0 V/common.
Signals are according to one of the Bronkhorst High-Tech B.V. standards. The model configuration contains
a code, describing the input/output signals.

Notes:
a. Max. load current output (sourcing)

: 375 Ohm

b. Min. load voltage ouput

: 10 kOhm

c. Input load resistance (voltage)

: 22K4

d. Input load resistance (current) (sinking)

: 200 Ohm

e. Input load resistance (current) (sourcing)

: 250 Ohm

1.4.3.2 Relay contacts

The (isolated) relay contacts are available at the following pins:
pin 5: common
pin 6: normally open
pin 7: normally closed

Note: contact arrangement during "power on"/"no alarm".
The specifications of the relay contacts are:
Contact arrangement

: SPCO (Single Pole Changeover)

Contact rating

: Switch voltage

: 24 Vac/dc max.

Switch current

: 0,2 A ac/dc max.

Switch power

: 3 W/3 VA max.


1.4.3.3 Reset input

The reset input signal should be connected to pin 8 (+ V) and pin 3 (0 V/common).
Reset can be achieved by either pulling the + V input (pin 8) to 0 V (pin 3) by means of a contact or a
transistor, or by applying an active low logic signal to pin 8 (+ V) and pin 3 (0 V). The logic signal level should
be 5 V (high) or 0 V (low).


1.4.4 Connection to FLOW-BUS

1.4.4.1 RJ45 connector

The shielded RJ45 modular jack connector has the following pin configuration:

pinnumber description
1
2
3
4
5
6
7
8

+ 15 Vdc supply
0 V (ref to +15Vdc )
Shield
0 V (ref to +15Vdc )
+ 15 Vdc supply
0 V (RS485)
RS485 - B
RS485 - A

1.4.5 RS232/FLOW-BUS interface

1.4.5.1 General description

The RS232/FLOW-BUS interface is an interface between the FLOW-BUS and the RS232 V24 serial
(computer) port.
It will either be supplied as a separate enclosed unit with a FLOW-BUS connector and a RS232 connector or
as an integral 14TE module of your E-7000 readout and control system.
The converter offers communication with a baudrate up to 38,4 kbaud.
Communication software support is available. Consult manual RS232/Flow-Bus interface.