Menu descriptions


4100 0000 Counter Reset Enable.

Here you can enter how the counter may be reset.

4110 0000 Remote manual reset. Here you can enable/disable a counter reset by the keyboard on the

instrument.

4120 0000 Remote external reset. Here you can enable/disable a counter reset by an external signal on the

instrument.

4130 0000 Remote FLOW-BUS reset. Here you can enable/disable a counter reset on the instrument via

the FLOW-BUS by means of f.i. a Personal Computer.

4140 0000 Remote auto reset. Here you can enable/disable a counter reset on the instrument when the

counter has reached its limit.

4150 0000 Local manual reset. Here you can enable/disable a counter reset by the keyboard on this

module.

4160 0000 Local external reset. Here you can enable/disable a counter reset by an external signal on this

module.


4200 0000 Counter Mode.

Here you can select how the counter must behave.

4210 0000 This option will disable the counter.
4220 0000 This option will let the counter count-up, but no limit will be checked.
4230 0000 This option will let the counter count-up, and check if a limit has been reached.

4300 0000 Edit Counter Unit.

Here you can change the unit for the counter to display the counter value.

4310 0000 Counter Unit Editor.

With UP/DOWN you can change the unit to be counted with.


4400 0000 Edit Counter Limit.

Here you can edit at which value the counter will reach its limit.

4410 0000 Counter Limit Editor.

With UP/DOWN you can change the number at the cursor, and with LEFT/RIGHT you go to
another cursor position.


4500 0000 Remote relay at limit.

Here you can enter what you want the instrument relay to do when a limit has been reached.

4510 0000 This option will disable a relay action at a counter limit.
4520 0000 This option will let the relay pulse at a counter limit.
4530 0000 This option will let the relay be continuously on at a counter limit.

4600 0000 Local relay at limit.

Here you can enter what you want the module relay to do when a limit has been reached.

4610 0000 This option will disable a relay action at a counter limit.
4520 0000 This option will let the relay pulse at a counter limit.
4530 0000 This option will let the relay be continuously on at a counter limit.

4700 0000 Limit Setpoint.

Here you can configure what to do with the setpoint at a limit. You can make the setpoint go to a
predefined value temporarily until reset situation.

4710 0000 This option wil not change the setpoint.
4720 0000 This option will change the setpoint to the entered value.
4721 0000 Setpoint At Counter Limit Editor.

Here you can enter the setpoint value to be active until reset.


NOTE:
Counter will be disabled when sensor type of instrument is set on either ‘Pressure’ or ‘Other’.