Menu descriptions

2100 0000 Operation mode. Here you can change the source of the setpoint for the operating instrument
2110 0000 This option selects a setpoint change with the keyboard or by means of a FLOW-BUS device

(operation module or Personal Computer).

2120 0000 This option selects a setpoint change coming from an analog input.
2130 0000 This option selects a setpoint change coming from an analog input, multiplied by a slave-factor

(analog master/slave).

2140 0000 This option selects a setpoint change by means of a FLOW-BUS device (operation module or

Personal Computer).

2150 0000 This option selects a setpoint change coming from an output signal of a master device, via

FLOW-BUS, multiplied by a slave-factor (FLOW-BUS master-slave).

2151 0000 Choose a master channel. With LEFT/RIGHT you can search for the master device on the

FLOW-BUS. Your instrument module will be slave.

Each time a possible master device has been found on the bus, the type and identification

number will be displayed. When ‘----‘ is displayed, this means that no master could be found.
With UP you can stop the search, and with 'enter' you select the currently displayed device.
Slave factor must be entered at the position where setpoint is normally entered

2160 0000 This option selects a setpoint change coming from an output signal of a master device, via

FLOW-BUS, multiplied by a slave-factor (FLOW-BUS master-slave), coming from an analog
input.

2161 0000 Choose a master channel. With LEFT/RIGHT you can search for the master device on the

FLOW-BUS. Your instrument module will be slave.

Each time a possible master device has been found on the bus, the type and identification

number will be displayed. When ‘----‘ is displayed, this means that no master could be found.
With UP you can stop the search, and with 'enter' you select the currently displayed device.
Slave factor must be entered at the position where setpoint is normally entered

2170 0000 This function sets a setpoint of 100%
2180 0000 This function sets a setpoint of 0%

2200 0000 Fluidset selection. Here you can change the actual fluidset. Your current settings will be saved

and the settings of the selected fluidset will become the new active settings for your instrument.
When fluidset changes, the instrument will use the calibration settings in the memory for sensor
signal/setpoint adjustments for that fluid.

2210 0000 Select the active fluidset by selecting a number from 1 to 8. Settings will be active after pressing

'enter'. This is for selection fluidsets only. For changing fluid calibration settings use Instrument

menu.

2300 0000 Setpoint Slope. Here you can enter a setpoint slope for this device. When the setpoint has been

changed and not overruled by counter limit or an alarm, the setpoint will (smoothly) increase to

the set value.

2310 0000 Edit the setpoint slope time.This is the time it will take for the setpoint value to increase from 0%

to 100% (and vice versa). The value can be entered from 0...3000.0 seconds.


2400 0000 Setpoint offset correction. Here you can enter a setpoint offset. This could be used to remove

cable offsets.

2411 0000 Edit setpoint zero offset. A value can be entered from -2.5%...+2.5%.