6 troubleshooting – Bronkhorst E-7000 User Manual

Page 66

Advertising
background image

BRONKHORST HIGH-TECH B.V.

page 66

9.17.004

6 TROUBLESHOOTING



LCD is not lit after 'power on'

check power input

check

fuse

Improper

reading

check module settings

No

reading/control

check instrument cabling

Improper system operation (alarm, counter)

check module settings

Instrument does not respond to setpoint given at
keyboard

setpoint source at operation menu should enable
setpoint at "keyb. & FLOW-BUS". (other settings
means different behaviour)

setpoint could have been overruled at an alarm
situation or counter limit (first perform reset)

setpoint slope has a large value. (setpoint will
slowly change in time) Change slope value if
necessary.

setpoint could be changed by other devices
connected to the FLOW-BUS.

check polynomial factors for setpoint and
measure at instrument menu.

check communication with FLOW-BUS.
If applicable

Setpoint menu cannot be reached

check setting of connected instrument type at
instrument menu. For sensor types setpoint
facilities will be disabled.

Counter/alarm is not responding to reset

check if reset-source has been enabled at reset
menu of reset counter/alarm.

Instrument seems to give inaccurate values

check if correct fluid set (with calibration factors)
has been selected (see fluidname).

Display shows ‘---‘

check if the operating instrument is active

check if the operating instrument is still at the
same address.

check instrument and module cabling

Other

problems

check instrument and operating module
parameters.


Advertising