Vi. tcx 1300 configuration and calibration – Digi-Star TCX-1000 SERIES User Manual

Page 24

Advertising
background image

D3556 05/29/03 REV. 2.2

Copyright 2001

24







VI. TCX 1300 CONFIGURATION AND CALIBRATION

A.

CONFIGURATION / CALIBRATION MENU

After pressing the internal calibration switch “SW1” on the circuit board directly above the
keyboard pigtail slot, the internal counts will be displayed. If the unit has not been calibrated
before or the e-prom has been replaced, the raw counts may appear to jump from one value to
another thousands of counts away. An auto-configuration may help solve this problem.
Once the raw counts are showing press the enter key once to display ConFig. Pressing ESC after
this will display the menu headings listed below.

199999 = Internal counts 000000 to 400000 under normal conditions, approximately 199999 with
no load cell connected. If there should ever be an error message displayed, pull the load cell
connector off and correct the problem.

ConFig
= Configuration Setup
SetPtS = Setpoint Setup
ACLtrS = Accumulation Entry
CAPCty = Capacity Entry
CAL 0 = Zero Calibration
CALPt1 = Span #1 Calibration, to
CALPt8 = Span #8 Calibration