SETPT/ALARM KEY

The set-point setup mode can only be entered from the weigh mode for which the set point is
configured.

Press the set-point key to display the set-point weight settings.

NOTE: Set-Points may only be setup in the primary (LB or kg) units, but are functional in both
the primary and alternate units.

If the value needs to be changed then set the new value per the following:

Press the NO/CLEAR key to clear the value to zero.

Press the YES/ENTER key to save the new value


NOTE: The set point responds to the absolute value of the weight reading (minus signs are
disregarded). Thus, a negative weight condition may be used to activate the set point for weigh
out applications.

K.

ANNUNCIATORS

Eight enunciators are located to the left and right of the upper display.

ID#

This annunciation is on in the id# entry/recall sequence.


MOTION

when on motion is present on the scale.


LB and KG

Either the LB or the kg annunciator will be on whenever the indicator is in the
one of the weigh modes.


TARE WEIGHT
On in the manual tare entry mode.

GROSS WEIGHT When on the indicator is in the gross weigh mode

NET WEIGHT When on the indictor is in the net weight mode

SET PT
On in the setpoint entry mode.

CENTER OF ZERO

A letter “c” displayed in the top half of the most significant digit.

When on, the gross weight is within ¼ of a graduation of true zero.

L.

ERROR MESSAGES

o-LoAd Whenever the weight on the scale exceeds 102.5% of the scale capacity the error message

o-LoAd will be displayed.


U-LoAd The scale may need servicing. Disconnect Load-Cell cable and press the word ZERO,

then reattach cable.


No bAL The printing was inhibited by scale motion. This message may be displayed during the

calibration procedure or during the printing process. Look at the raw counts and see if