D3556 05/29/03 REV. 2.2

Copyright 2001

27

Note: The preceding steps have been provided to adjust the indicator for the best response time
and the best vibration tolerance. The Filter Readings sets the maximum number of internal
readings that may be averaged together to smooth out vibration and other scale fluctuations. The
TCX 1300 takes a reading 16 times per second, thus the filter time is adjustable from .125 seconds
to 2 seconds. The Filter Off Readings sets the number of consecutive readings that differ more
than this amount that will be ignored; the next such reading will end filtering. The majority of
applications where the scale never settles, such as a cattle scale, the High Vibration filter will need
to be selected. In cases where very small changes in weight must be displayed quickly, the filter
readings may be adjusted downward. In case where “spike” readings are common, the filter off
readings may be adjusted upwards.

CONTROL DISPLAYED PROGRAM VALUE

AUTO

= 1

AUTO =

2

AUTO =

3

Display Rate

dSp=16

16 updates per sec

16

16

4

dSp=8

8 updates per sec

dSp=4

4 updates per sec

dSp=2

2 updates per sec

dSp=1

1 update per sec

Auto Zero
Tracking Readings

A0rd00 AZT

Disabled

40

40 40

A0rdxx

xx = 01 to 99 : Number of
readings within AZT window
to actuvate zero

Auto Zero
Tracking Window

A0gd.5 0.5

Graduations

3 1 .5

A0gd01 1

Graduations

A0gd02 2

Graduations

A0gd03 3

Graduations

Stability Detection
Readings

Sbrd00 Motion

Disabled

10

10 10

Sbrdxx

xx = 01 to 99 : Number of
readings in stability window
to deactivate motion

Stability Detection
Window

Sbgd0.5 .5

Graduations

3 1 3

Sbgd01 1

Graduations

Sbgd02 2

Graduations

Sbgd03 3

Graduations

Keyed Zero Band

0b=1.00

100% of capacity

0.02

0.02

1.0

0b=0.02

2% Of Capacity this option is
set to .02 it the scale is
involved in NTEP weighing.
This is used to discourage
illegal use of the scale system

Units Conversion

1 Unit

Calibration Unit Only

1

1

1

2 Unit

Conversion Enabled. This
option will allow a conversion
from one set to the other I the
2 unit setting

Net Mode Select

nEtoFF

Net Mode Disables

on

on

On

nEt on

Net Mode Enabled

Auto Log

AtE=on

Automatically logging loading
or unloading weights, once
reaching or exceeding
tolerance setting

Off Off Off