Stockweigh 600 – Digi-Star StockWeigh 600 User Manual
Page 14
StockWeigh 600
10
.
+
The Short Form Setup & Calibration procedure allows you to change
the “SETUP” and “CAL” numbers of the indicator.
Following is a list of functions that are controlled by the “SETUP”
number:
Weigh Method (
)
Display Units (
*
)
Display Counts (
)
Gain
Scale Capacity
The “CAL” number is adjusted to make the scale read the proper
weight for different load cells and to make accuracy adjustments
on a scale system. Systems should be checked and adjusted if
necessary using known weights to insure accuracy.
!/
! (
#
0, ,
1
Before continuing, first write down the current SETUP and CAL
numbers of your StockWeigh indicator. These numbers are displayed
during the Self Test. To run the self-test with the indicator already ON,
press [Net/Gross], then [On/Off] to start the Self Test. Press [On/Off] to
"pause" the Self-Test while numbers are displayed. Press [On/Off] again
to "resume"
SETUP # _______________ CAL # ________________
To accurately calibrate the scale, you will need a large amount of weight
that has a known value. For best results you should have at least as
much weight as the largest load you plan to weigh.
,
2
' % !
,/ $
Zero-Balance the scale so the display reads zero.
Put the KNOWN WEIGHT on the scale platform and write down the
WEIGHT DISPLAY.
Perform the following equation to find the ACCURATE CAL #.