Stockweigh 600 – Digi-Star StockWeigh 600 User Manual

Page 14

Advertising
background image

StockWeigh 600

10

.

+

The Short Form Setup & Calibration procedure allows you to change

the “SETUP” and “CAL” numbers of the indicator.

Following is a list of functions that are controlled by the “SETUP”

number:

Weigh Method (

)

Display Units (

*

)

Display Counts (

)

Gain
Scale Capacity

The “CAL” number is adjusted to make the scale read the proper

weight for different load cells and to make accuracy adjustments

on a scale system. Systems should be checked and adjusted if

necessary using known weights to insure accuracy.

!/

! (

#

0, ,

1

Before continuing, first write down the current SETUP and CAL

numbers of your StockWeigh indicator. These numbers are displayed

during the Self Test. To run the self-test with the indicator already ON,

press [Net/Gross], then [On/Off] to start the Self Test. Press [On/Off] to

"pause" the Self-Test while numbers are displayed. Press [On/Off] again

to "resume"

SETUP # _______________ CAL # ________________
To accurately calibrate the scale, you will need a large amount of weight

that has a known value. For best results you should have at least as

much weight as the largest load you plan to weigh.

,

2

' % !

,/ $

Zero-Balance the scale so the display reads zero.
Put the KNOWN WEIGHT on the scale platform and write down the

WEIGHT DISPLAY.
Perform the following equation to find the ACCURATE CAL #.