3 remote reset input wiring – Tecfluid MX4 For FLOMAT Series User Manual

Page 18

Advertising
background image

18

Terminal

Description

Relay

6

Normally

Open

Relay

1

5

Common

Relay

1

4

Normally

Open

Relay

2

3

Common

Relay

2

The relay outputs provide relays with potential free contacts (see characteristics in page
50).

The status of the relay contacts corresponds to the relay at rest

.

The relay contacts are not protected in any way, and therefore they must be installed
externally

as required in the application

, taking into account the limitations of the

characteristics of such contacts. In the case of having inductive loads, and to extend the
working life of the relay contacts, it is recommended to use overvoltage protection (VDR
for AC and diodes for DC loads). In all cases a fuse or some kind of protection against
short circuits,

should always be provided

according to the needs of the intended load.

2.2.3

Remote reset input wiring

2

INSTALLATION OF THE ELECTRONIC CONVERTER