3 analog output, 1 4-20 ma output configuration, 2 programming of the 4-20 ma output – Tecfluid MX4 For FLOMAT Series User Manual

Page 40: 3 current calibration for 4 and 20 ma

Advertising
background image

40

7.5.3.1 4-20 mA output configuration

It allows to choose between unidirectional and bidirectional.

7.5.3 Analog

output

The flow rate corresponding to 4 mA and 20 mA can be programmed and the calibration

for these two values can be made as well.

Programming a flow rate value for 20 mA higher than the one for 4 mA, the loop current
changes from 4 to 20 mA as the flow increases.

Otherwise, programming a flow rate value for 20 mA lower than the one for 4 mA, the
current loop will change from 20 to 4 mA as the flow increases.

7.5.3.3 Current calibration for 4 and 20 mA

The MX4 converter is delivered with the current output already calibrated. If you want to
correct a derive of the 4 or 20 mA current values because they do not coincide with the
ammeter used, it can be done by pressing the keys (Down / Left) and (Up) to increase or
decrease the current respectively in each case. The values shown on the screen are
reference levels. When one of this data is confirmed, the MX4 will take the current value
shown on the ammeter as a reference.

If “Unidirectional” is chosen, the analog output will have a value proportional to the flow
rate when positive, and it will provide always 4 mA when negative.

If “Bidirectional” is chosen, the analog output will have a value proportional to the flow rate
both for positive and negative values.

7.5.3.2 Programming of the 4-20 mA output

Flow rate values equivalent to each current are entered directly.

7

CONVERTER PROGRAMMING