3.2 Cable

installation

The following points must be taken into account:



The cable should be installed in a conduit or should be securely fixed, given that
movements of the cable can induce reading errors.



The cable should be placed as far as possible from sources of electrical noise such as
switching gear and electrical machines.



The maximum length of the connection cable depends on the fluid conductivity. For
liquids with a conductivity higher than 500

µS/cm cable length can be up to 150 m.

3.3

Cable connection to sensor

For the sensor connection, first loosen the cable gland to allow the cable to slide in it.



Make sure that the two fixing screws of the cover do not protrude inside the cover (this
avoids damaging the o-ring of the sensor neck).



Apply a few Vaseline on the sensor neck to aid inserting in the cover, specially on the o-
rings.

Connector cover

Mail connectors

Sensor neck

Fixing screws

3

REMOTE SENSOR