Page 42

FSS Series Fountain Solution Systems

IAE-600

5

Setup Procedures

5-1 Conductivity Meter

GLI PRO-SERIES

1) Press the menu key.
2) Scroll down to configure and press enter.
3)
Scroll down to set C (Celsius) or F (Farenheight) and press

enter.

4) Scroll up to change C (Celsius) to F (Farenheight) and press

enter.

5) Scroll down to sensor and press enter.
6) Scroll down to display format and press enter.
7) The screen should read display format.
8)
Press the arrow up or down until screen reads <20.00 mS/CM>

and press enter.

9) Scroll down to temp element and press enter.
10)
Scroll down to Set “T” factor? and press enter.
11)
Look at the label on the sensor to get the resistance value. (eg:

T=1000.5)

12) Enter the value found on the sensor and press enter.
13)
Scroll down to exit then press enter. (This will take you back

to the sensor screen)

14) Scroll down to exit again and press enter. (This will take you

back to the configure screen)

15) Scroll down to set output and press enter.
16)
Scroll down to 4 mA value and press enter.
17)
Change to 0.00 mS/CM and press enter.
18)
Scroll down to 20.00 mA and press enter.
19)
Set the 20 mA value to 5.00 mS/CM and press enter.
20)
Scroll down to exit then press enter. (This will take you back

to the configure screen)

21) Scroll down to exit and press enter. (This will take you back

to the main menu screen)

22) Scroll down to calibrate then press enter.
23) Sensor
should appear then press enter.
24)
Scroll down to zero then press enter.