Instruction Manual

760004-A
February 2002

P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

WARNING

.

POSSIBLE EXPLOSION HAZARD

Ensure that all gas connections are made as labeled and are leak free. Improper gas connections
could result in explosion or death.

CAUTION

.

OVER-VOLTAGE SPIKING

If this analyzer module is used with a non-Rosemount Analytical power supply, adding Rosemount
Analytical PN 903341 Current Protector in series with the 24V positive power line will prevent over-
voltage spiking and resultant fuse blowing when powering up the instrument.

CAUTION

.

PRESSURIZED ENCLOSURE

This enclosure shall not be opened unless the area is known to be free of flammable materials or
unless all devices within have been de-energized.

Area classification for the protected enclosure:

Nonclassified.

Pressurization: Type Z

Temperature Identification Number: T4A

Power shall not be restored after enclosure has been opened (or loss of purge) until enclosure has
been purged for a minimum of 6 (six) minutes at the minimum pressure of 689 hPa (10 psig).