Maintenance and Service 4-13

Model NGA 2000 HFID

b. Cartridge

Heater

1. Remove oven from analyzer module

per Section 4-2a.

2. Remove burner/thermal block from

oven per Section 4-2b.

3. Refer to Figure 4-9. Loosen retaining

set screw, pull out cartridge heater.

4. Install replacement cartridge heater,

snug down set screw.

5. Install burner/thermal block into oven.

6. Install oven into analyzer module.

c. Thermostat

1. Remove oven from analyzer module

per Section 4-2a

2. Remove burner/thermal block from

oven per Section 4-2b.

3. Refer to Figure 4-9. Remove the two

retaining screws, pull thermostat out.

4. Install replacement thermostat, attach

with the two retaining screws.

5. Install burner/thermal block into oven.

6. Install oven into analyzer module.

d. Sample

Capillary

1. Remove oven from analyzer module

per Section 4-2a

2. Remove burner/thermal block from

oven per Section 4-2b.

3. Remove burner from thermal block

per Section 4-4a

4. Refer to Figure 4-10. Remove the

two screws securing the capillary
cover to thermal block, remove cover.

5. Remove capillary nut, remove capil-

lary.

6. Install replacement capillary.

7. Insert capillary into thermal block.

The capillary may require bending to
fit.

8. Install

cover.

Figure 4-10. Thermal Block Assembly

Capillary

Cover