Maintenance and Service 4-3

Model NGA 2000 HFID

Oven Assembly
659551

4-2 OVEN

Though the oven can be replaced as a com-
plete unit, all internal components are field re-
placeable.

a. Removal

Refer to Figure 4-3, disconnect the oven's
three gas lines and seven electrical ca-
bles, noting location of mating connectors
for re-installation.

NOTE

DO NOT remove the fittings from the
gas lines on the detector.

Remove the two hex nuts securing the
oven to the chassis and the two
screws securing oven to the rear
panel. Lift oven assembly from ana-
lyzer.