A. lon/power module, B. led indicator assembly, C. manual ignite toggle switch – Emerson Process Management 760004-A User Manual

Page 63: D. burner air sensor, E. fuel sensor, F. burner air and fuel regulators, Model nga 2000 hfid

Advertising
background image

Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-25

Model NGA 2000 HFID

a. LON/Power

Module

1. Disconnect wiring connectors, note

locations.

2. Refer to Figure 4-20. From the out-

side of the front panel, remove the
two mounting screws.

3. Install replacement module in reverse

order.

b. LED

Indicator

Assembly

1. Disconnect wiring connector, note lo-

cation.

2. Refer to Figure 4-20. From the inside

of the front panel, remove the two hex
nuts securing LED indicator assembly
to front panel. Remove indicator as-
sembly and o-rings (four).

3. Inspect o-rings for damage, replace if

necessary. Install o-rings on re-
placement indicator assembly, mount
assembly on mounting studs with hex
nuts.

4. Re-connect

wiring

connector.

c. Manual Ignite Toggle Switch

1. Disconnect wiring connector, note lo-

cation.

2. Refer to Figure 4-20. From the out-

side of the front panel, remove the
toggle switch seal.

3. Pull the switch and o-ring out from in-

side the front panel.

4. Inspect o-ring for damage, replace if

necessary. Install o-ring on replace-
ment switch, insert through front
panel from the inside.

5. Install switch seal.

6. Re-connect

wiring

connector.

d. Burner Air Sensor

1. Disconnect wiring connector, note lo-

cation.

2. Using an open-end wrench to hold the

sensor fitting while using another
open-end wrench to remove the sen-
sor.

3. Replace the Teflon pipe thread tape

on the treads of the sensor fitting.

4. Install sensor onto sensor fitting.

5. Re-connect

wiring

connector.

e. Fuel

Sensor

1. Disconnect wiring connector, note lo-

cation.

2. Using an open-end wrench to hold the

sensor fitting while using another
open-end wrench to remove the sen-
sor.

3. Replace the Teflon pipe thread tape

on the treads of the sensor fitting.

4. Install sensor onto sensor fitting.

5. Re-connect

wiring

connector.

f.

Burner Air and Fuel Regulators

1. Disconnect the two tubes and the

sensor fitting on the rear of the regu-
lator, note locations.

2. Replace the Teflon pipe thread tape

on the threads of the sensor fitting.

3. Remove the regulator and o-ring.

4. The replacement regulator comes

with two panel mounting nuts, remove
both and discard one of them.

Advertising