4-16 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

b. Case Temperature Sensor

1. Per Section 4-6, remove securing

hardware from electronics assembly
and lay on bench.

2. Disconnect case temperature sensor

cable.

3. Remove screw securing cable clamp

holder to signal board.

4. Remove case temperature sensor

from cable clamp holder.

5. Per Figure 4-13 insert replacement

case temperature sensor into cable
clamp holder.

6. Re-assemble to signal board mount-

ing screw.

Figure 4-13. Case Sensor Installation

Center sen-

sor in cable