C. burner jet installation, Model nga 2000 hfid – Emerson Process Management 760004-A User Manual

Page 49

Advertising
background image

Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-11

Model NGA 2000 HFID

c. Burner Jet Installation

WARNING

BURNER CONTAMINATION

Do not handle internal parts of the burner
with bare hands. All tools used for main-
tenance must be free of contaminates.

1. Install

new lower gasket, jet assembly

and upper gasket into burner base,
finger-tight jet nut.

2. Install new sample jet (with gasket)

and tighten.

3. Tighten jet nut.

4. Install air baffle per Figure 4-8.

NOTE

Incorrect installation of air baffle will
cause ignition failure.

5. See Figure 4-7B. Insert new o-ring

into burner base.

6. Set combustion chamber into burner

base being careful not to move air
baffle
.

7. See Figure 4-7A. Insert new gasket

on combustion chamber, install
burner cap and burner cap retainer,
torque screws to 6 inch lbs.