Instruction Manual

760004-A
February 2002

4-6 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

4-3 BURNER

This section covers burner components which
can be replaced without removal of oven from
the chassis.

a. Temperature

Sensor

1. Refer to Figure 4-4A. Remove the

four screws on the oven cover, re-
move cover.

2. Refer to Figure 4-5. Remove the

burner cap retainer.

3. Disconnect the temperature sensor

wiring connector, note location.

4. Remove the temperature sensor.

5. Insert replacement sensor.

NOTE

The leads of the temperature sensor
must be leading away and down from
the sensor to enable proper fit of
burner cap retainer.

6. Install the burner cap retainer. U-slot

must be located above temperature
sensor.

7. Re-attach

wiring

connector.

8. Install oven cover.

b. RTD

Detector

1. Refer to Figure 4-4A. Remove the

four screws on the oven cover, re-
move cover.

2. Refer to Figure 4-5. Loosen the set

screw securing RTD detector.

3. Disconnect RTD detector wiring con-

nector, note location.

4. Gently grasp RTD detector wires and

pull out of hole.

5. Insert replacement RTD detector into

hole, snug down set screw.

6. Re-attach

wiring

connector.

7. Install oven cover.

c. Igniter

1. Refer to Figure 4-4A. Remove the

four screws on the oven cover, re-
move cover.

2. Refer to Figure 4-5. Disconnect the

igniter wiring connector, note location.

3. Using an open-end wrench, unscrew

the igniter assembly from the burner.
Verify that o-ring is also removed.

4. Install replacement igniter and new o-

ring. Using open-end wrench, snug
down. Do not over-tighten!

5. Re-attach

wiring

connector.

6. Install oven cover

d. Flameout

Sensor

1. Refer to Figure 4-4A. Remove the

four screws on the oven cover, re-
move cover.

2. Refer to Figure 4-5. Disconnect the

flameout detector wiring connector,
note location.

3. Lift up the burner cap until flameout

sensor is accessible. Using an open-
end wrench, unscrew the flameout
detector from the burner. Verify that
o-ring is also removed.

4. Install replacement flameout detector

and new o-ring. Using open-end
wrench, snug down. Do not over-
tighten!

5. Re-attach

wiring

connector.

6. Install oven cover.