C. gas requirements, Model nga 2000 hfid, Instruction manual – Emerson Process Management 760004-A User Manual

Page 19

Advertising
background image

Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5

Model NGA 2000 HFID

c. Gas

Requirements

Sample .......................................... Non-flammable, below 100% of LEL

Flow rate ................................ 1.0 to 2.5 L/min.
Supply pressure..................... 345 to 620 hPa-gauge (5 to 9 psig)
Temperature .......................... 110

o

C to 230

o

C (230

o

F to 446

o

F), <20

o

C variance/24 hours, <10

o

C

variance/hr.

Particulates............................ filtered to <2 microns
Dewpoint................................ <15

o

C below the setpoint

Regulated Air................................. Instrument air or nitrogen

Flow rate ................................ 2 to 4 L/min.
THC .......................................

≤2 ppm, CH

4

Supply pressure..................... 689 to 1723 hPa-gauge (10 to 25 psig)
Particulates............................ filtered to <2 microns

Purge Air:....................................... Instrument air, nitrogen or other nonflammable gas (refer to

ANSI/NFPA 496 for the requirements for the Protective Gas Sys-
tem)

Flow rate: .............................. 16 to 18 L/min.
Supply pressure:.................... 689 to 1378 hPa-gauge (10 to 20 psig)

Fuel Gas ........................................ Premixed 40% hydrogen and 60% helium

Flow rate ................................ 80 to 100 ml/min
THC .......................................

≤0.5 ppm, CH

4

Supply pressure..................... 3101 to 3446 hPa-gauge (45 to 50 psig)

WARNING

EXPLOSION HAZARD

Do not use pure hydrogen fuel. An explosion resulting in severe personal injury or death could oc-
cur.

Burner Air ...................................... Zero-grade air

Flow rate ................................ 350 to 400 mL/min.
THC .......................................

≤1 ppm, CH

4

Supply pressure..................... 1723 to 3446 hPa-gauge (25 to 50 psig)

Advertising