Instruction Manual

760004-A
February 2002

5-2 Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

18. Verify that there is +10VDC to the

reference thermistor.

c. Drift

1. Verify that the sample, burner air, and

fuel supply pressures are constant.

2. Check that the tubing, regulators,

pumps, fittings, and valves are clean
of hydrocarbons.

3. Verify that the oxygen level in the

burner air and sample are constant.

4. Verify the THC level is correct for the

burner air and fuel supply.

5. Check that the ambient temperature

is changing <10

°C per hour.

6. Verify the burner is clean.

7. Verify temperature of the sample gas,

case, burner, and oven has stabilized.

8. Verify the Preamp PCB is clean.

9. Verify atmospheric pressure at burner

exhaust is constant.

10. Verify purge gas pressure is constant.

11. Verify burner has been on and stabi-

lized.

12. Check for gas leaks.

d. Noise

1. Check that the burner exhaust is free

from water condensation.

2. Verify connection to the collector is

correct.

3. Verify connection to the polarizing

voltage is correct.

4. Check the ambient temperature is

changing <10

°C per hour.

5. Verify the +24VDC is clean and

grounded properly.

6. Verify there are no strong magnetic

fields near.

7. Check for excessive vibration.

8. Verify burner exhaust is vented to a

constant atmospheric pressure.

9. Verify bypass line is vented to a con-

stant atmospheric pressure.

10. Verify purge out port vented to a con-

stant atmospheric pressure.

11. Verify the collector wires are routed

away from the heater.

12. Verify the collector wires are clean

and not damaged.