Instruction Manual

760004-A
February 2002

4-4 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

b. Disassembly

1. Refer to Figure 4-4A. Remove the

four retaining screws on the oven
cover, remove cover.

2. Remove the two screws and one nut

securing the outer oven front panel
to the outer oven, remove front
panel.

3. Remove the nuts and ferrules from

sample in and sample bypass out.

CAUTION

PREAMP CONNECTORS

The electrical preamp connectors are frag-
ile, handle with care to avoid breaking sol-
der connection.

4. Refer to Figure 4-4B. Remove the

two nuts and washers from the elec-
trical preamp connectors on the inner
oven front panel. Do not unsolder
these connections
.

5. Unscrew the three screws from inner

front panel and remove it.

6. Refer to Figure 4-4C. Remove the

two hex nuts securing the burner to
the bottom of the inner oven.

7. Disconnect the sample input and

output bypass fittings.

8. Lift the burner/thermal block up and

out, while disconnecting exhaust.

Reverse procedure for installation.