Instruction Manual

760004-A
February 2002

4-26 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

5. Inspect o-ring for damage, replace if

necessary. Install o-ring onto regu-
lator threaded shaft.

6. Insert regulator into front panel, se-

cure with mounting nut.

7. Re-attach the three tubes.

g. Purge Air Regulator

1. Remove the regulator mounting nut,

remove mounting bracket.

2. Loosen nut on tee fitting attached to

purge air flow switch.

3. Disconnect tube at elbow, remove

regulator.

4. Remove the two plugs, elbow and

male adapter fittings from the regula-
tor.

5. Replace the Teflon pipe thread tape

on the two plugs, the elbow and the
male adapter and install into replace-
ment regulator.

6. Connect tube to elbow, insert male

adapter into tee fitting.

7. Install mounting bracket onto regula-

tor, hand snug mounting nut.

8. Attach mounting bracket to front

panel, tighten regulator mounting nut.

h. Purge Air Flow Switch and Diffuser

1. Unscrew flow switch from tee fitting.

2. Replace Teflon pipe thread tape on

tee fitting.

3. Remove diffuser from flow switch and

install into replacement flow switch.

4. Install replacement flow switch.

5. Install purge switch onto tee fitting.

6. Re-connect

tubes.

i.

Burner Air Solenoid Valve

1. Disconnect the tube at the top elbow

fitting.

2. Disconnect the tube at the tee fitting,

remove valve analyzer module.

3. Holding the air ignite restrictor, un-

screw the solenoid valve.

4. On the solenoid valve, remove the

connector fitting.

5. Replace the Teflon pipe thread tape

on the elbow, connector and restric-
tor.

6. Verify replacement solenoid valve

wires (flat side of body) are exiting on
the same side as the COM port as
shown in Figure 4-20. If not, use an
open-end wrench to hold the N.O. hex
port while rotating body.

7. Install air ignite restrictor into N.C.

port.

8. Install elbow into COM port and con-

nector fitting into N.O. port.

9. Re-connect

tubes.

j.

Air Ignite Restrictor

1. On the burner air solenoid valve:

a. Disconnect the tube at the top el-

bow fitting.

b. Disconnect tube at tee fitting.
c. Lift solenoid valve from analyzer

module.

d. Disconnect tube going to air ignite

restrictor.

e. Remove restrictor from solenoid

valve.

2. Add Teflon pipe thread tape to re-

placement restrictor, install into
solenoid.

3. Re-connect tubes to restrictor, elbow

and tee fitting.