Instruction Manual

760004-A

February 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Maintenance and Service 4-15

Model NGA 2000 HFID

Figure 4-12. Electronics Assembly – Exploded View

a. Printed Circuit Boards

When replacing a circuit board, the fol-
lowing procedure is recommended:

Per Section 4-6, remove securing hard-
ware from electronics assembly and lay
on bench.

Remove securing hardware from printed
circuit board to be replaced, do not dis-
connect cable(s).

One at a time, remove the wiring con-
nectors and attach to replacement board.

Mount replacement board to electronics
assembly.

Safety Board
657499

Sensor Board
657060

Computer Analysis Board
659149

Preamp Assembly
656945

Case Pressure Purge
Switch 903690

Power Supply Board
655764

Case Temperature Sensor
656026