5 thermal block, A. sample rtd, Model nga 2000 hfid – Emerson Process Management 760004-A User Manual

Page 50

Advertising
background image

Instruction Manual

760004-A
February 2002

4-12 Maintenance and Service

Rosemount Analytical Inc. A Division of Emerson Process Management

Model NGA 2000 HFID

Sample RTD
657061

Cartridge Heater
659643

Thermostat
657065

4-5 THERMAL BLOCK

The sample RTD can be replaced with the
thermal block attached to burner and mounted
in oven. The cartridge heater and thermostat
are also replaceable with thermal block se-
cured to burner, but must be removed from
the oven.

a. Sample

RTD

1. Refer to Figure 4-4A. Remove the

four screws securing the oven cover,
remove cover.

2. Disconnect the sample RTD wiring

connector, note location.

3. Refer to Figure 4-9. Remove the two

screws securing the sample RTD, pull
sample RTD out.

4. Install replacement sample RTD, se-

cure with screws.

5. Attach sample RTD wiring connector.

6. Re-attach oven cover.

Figure 4-9. Thermal Block – Sample RTD, Cartridge Heater and Thermostat