C. electrical, D. sensors, Model 7003m – Emerson 7003M User Manual

Page 17

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Description and Specifications 1-5

Model 7003M

c. Electrical

Power Requirements..................... 115/230 VAC

±10%, 50/60 Hz ±3 Hz

Display........................................... 3-1/2 digit liquid crystal (LCD)
Signal Output................................. Isolated 4 to 20 or 0 to 20 mA (selectable)

600 ohms maximum load

Alarm Relay Contacts.................... Two Form C Relays

SPDT, resistive load, independently adjustable
Deadband adjustable over full range
Alarm Relay Contacts Ratings
General Purpose (Ordinary) Locations:
115 VAC, 3.0 A (resistive), 230 VAC, 1.5 A resistive
Class I, Division 2 Locations:
CSA: 120 VAC, 3.0 A (non-inductive)
240 VAC, 1.5 A (non-inductive).
FM: 28 VDC (resistive) (Ci = 0, Li = 0)
Groups A&B: 150 mA
Group C: 400 mA
Group D: 540 mA

Recorder Output............................ Isolated 4 to 20 mA

d. Sensors

Types............................................. Rechargeable and Disposable
Process Connections .................... Refer to flow chamber (if supplied)
Sample Pressure........................... 0 to 50 psig (0 to 345 kPa)
Sample Temperature..................... 32 to 110

°F (0 to 44°C)

Ambient Temperature.................... 32 to 110

°F (0 to 44°C)

Ambient Humidity .......................... 95% relative humidity, non-condensing (maximum)

Also refer to instructions supplied with sensor.