Routine Servicing and Troubleshooting 6-1

Model 7003M

SECTION 6

ROUTINE SERVICING AND TROUBLESHOOTING

WARNING

ELECTRICAL SHOCK HAZARD

Servicing requires access to live parts
which can cause death or serious injury.
Refer servicing to qualified personnel.

Relay contacts wired to separate power
sources must be disconnected before
servicing.

6-1 ROUTINE SERVICING

Most service and maintenance involves
cleaning and recharging the sensor. Refer
to the instructions supplied with the sensor
for recharging and maintenance procedures.

Service is rarely required on the Model
7003M Percent Oxygen Analyzer. Repair is
limited to replacement of PCB boards,
EPROM and fuses. See Section 7
Replacement Parts.

6-2 TROUBLESHOOTING

Troubleshooting should be referred to
qualified service technicians. Refer to
Section 5-3 Circuit Descriptions, on page 5-
2, as an aide in isolating a problem.

a. Symptoms

Calibration error code displayed

Possible cause
Exhausted Sensor

Check/Test
Install know good sensor and isolate

Corrective Action
Recharge or replace sensor

Cannot adjust signal to calibration
value

Possible cause
Sensor voltage error

Check/Test
Install know good sensor and verify
voltage

Corrective Action
Recharge or replace sensor

Readings not consistent with calibra-
tion curve

Possible cause
Incorrect sensor used for application

Check/Test
Verify sensor corresponds with
application

Corrective Action
Install new sensor and calibrate with
span gas

Possible cause
Sensor temperature incorrect

Check/Test
Sensor temperature should be 0 to 50

°C

Corrective Action
Install new sensor and calibrate with
span gas

Possible cause
Incorrect span gas concentration

Check/Test
Verify span gas concentration

Corrective Action
Install new sensor and calibrate with
span gas