June 2002

3-6 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

c. Error

Messages

If an error condition occurs, E-# (non-fatal
error) or F-# (fatal error) will flash on the

display. Table 3-10 through Table 3-13
list error codes and corrective actions.
Pressing "ACK" or "ENTER" will
acknowledge any alarm and the error
message will disappear.

CODE

DEFINITION

F-1

Invalid PROM (16#C000#..16#FFFF#) checksum detected

F-2

Internal RAM (16#00#..16#FF#) failed testing

F-3

External RAM (16#4000#..16#5FFF) failed testing

F-4

EEPROM (16#B600#.. 16#B7FF#) bulk erase failure

F-5

Signal Board not connected

F-6 thru F-9

EEPROM write failure

Table 3-10. Power ON, Reset Diagnostics Errors (Fatal Errors

CODE

DEFINITION

E-1

5 volts power supply above 5.5 V

E-2

5 volts power supply below 4.5 V

E-3

12 volts power supply above 13.2 V

E-4

12 volts power supply below 10.8 V

E-5

sensor output too low for calibration

E-6

sensor missing

E-8

sample temperature is below 1

°C

E-9

sample temperature is above 44

°C

Table 3-11. RUN Mode Errors (Non-Fatal Errors

CODE

DEFINITION

HI

sample temperature is above 50

°C

LO

sample temperature is below 0

°C

Ahi

excessive sensor signal

Table 3-12. Signal Board Errors

CODE

DEFINITION

COP

Computer Operating Properly error (error detected)

iot

Invalid Op code Trap (internal computer command is invalid)

E-0

power interruption occurred

Table 3-13. Microprocessor Errors That Reset The System