Model 7003m, Operator manual – Emerson 7003M User Manual

Page 52

Advertising
background image

Operator Manual

748223-K
June 2002

4-6 Calibration and Adjustments

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

CARET

LIT

VALUE DISPLAYED

TYPICAL

VALUE

1

ADJUSTABLE

%

Raw Sensor Voltage

2

Yes

°C

Thermistor Voltage

2.43

No

-

3

Current Output

17.43

Yes

Use ARROW keys to make adjustments.

Table 4-7.

Caret Readings

1

These typical values were obtained under the following conditions:

O2 Concentration:

20.95%

Sensor Temperature:

26.07

°C

Current Output Range:

4-20mA

Range Endpoint, Low:

0%

Range Endpoint, High:

25%

Zero Offset Signal:

None

Pressure Compensation:

None

2

Typical voltages for new sensors at 25

°C:

TYPE OF SENSOR

MINIMUM

NOMINAL

MAXIMUM

Disposable

1.05V

1.39V

1.69V

Rechargeable

0.97V

1.29V

1.76V

Near the end of the life of the sensor, the sensor voltage is typically 0.53V (disposable) or 0.48V (rechargeable). At tempera-

tures (T) other than 25

°C, multiply the voltages above by (1.025)exp(T-25).

3

"-" is a minus sign displayed on the left of the display between AL 1 and AL 2.