3 hold mode, Model 7003m – Emerson 7003M User Manual

Page 51

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Calibration and Adjustments 4-5

Model 7003M

4-3 HOLD MODE

Refer to Figure 4-2 on page 4-7 and Table 4-6
below. In the HOLD mode:

All instrument functions are locked

All existing values are retained

The instrument stops responding to
sensor signals

The last measured value is displayed

The last output value remains at the
isolated current output connections

Alarms and ON/OFF-Controllers are held
in their last state

No keys are recognized except entry to
the diagnostic routing.

KEY

DISPLAY

WHAT HAPPENS

DISPLAY

Nothing

Nothing

ACK

Nothing

Nothing

ALARM 1

Nothing

Nothing

ALARM 2

Nothing

Nothing

UP ARROW

Nothing

Nothing

DOWN ARROW

Nothing

Nothing

CAL

Nothing

Nothing

ENTER

"run"

Drops back to RUN mode

DISPLAY + ACK

"d-4" "tSt"

Moves to DIAGNOSTIC mode

Table 4-6.

HOLD Mode Key Functions

If the sensor cable must be disconnected, the
instrument can be put in the HOLD mode, in
which all existing signal levels and output
current levels and output current levels are
maintained. This may be useful to prevent
disturbing an ON/OFF-control loop connected
to the relay contacts.

1. Press the UP ARROW and DOWN

ARROW at the same time to put the
instrument on HOLD.

2. If the security is activated, enter the

security identification (id) code and press
ENTER to continue.

3. Hld, YES, flash briefly, then Hld flashes

alternately with the last oxygen reading.
The output signal will remain unchanged
until the instrument returns to the RUN
mode.

4. Press ENTER to return to the RUN mode,

or continue with Section 4-4 Diagnostics
Mode on page 4-8.