background image

Operator Manual

748223-K
June 2002

3-10 Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

3-4 RUN MODE

Refer to Figure 3-2 on page 3-2 and Table
3-15 below. The RUN mode is the normal
mode of operation, from which the operator
can:

• Display oxygen concentration and sensor

temperature

• Acknowledge

alarms

• Display security code
• Access other modes

KEY

DISPLAY

WHAT HAPPENS

DISPLAY

Number change

Toggles between O

2

reading and temperature

ACK

Clears message

Acknowledges Alarms

ALARM 1

"A-1" then ON

To Alarm 1 SETUP mode

ALARM 2

"A-2" then ON

To Alarm 2 SETUP mode

UP ARROW

"SrY" "run"

Stays in RUN mode

DOWN ARROW

"SrY" "run"

Stays in RUN mode

CAL

"CAL" then ON

To CALIBRATION mode

ENTER

"SrY" "run"

Stays in RUN mode

CAL + ENTER

"SEt" then ON

Moves to SETUP mode

UP + DOWN

"hLd" then ON

Moves to HOLD mode

DISPLAY + ACK

"123"

Shows the security code

Table 3-15. RUN Mode Key Functions

a. Display Oxygen Concentration/Sensor

Temperature

Press DISPLAY once to display sensor
temperature. Press DISPLAY again to
display oxygen concentration.

b. Acknowledge

Alarms

If an alarm setpoint has been reached, "A-
1" or "A-2" will flash with a concentration
value. Acknowledging the alarm will stop
the flashing display and turn off the relay
switch.

To Acknowledge Alarm - Press ACK
key. If relays are configured as ON/OFF
controllers, pressing the ACK key has no
effect.

c. Access Other Modes

The following lists which key(s) to press to
access various modes:

SETUP

= CAL+ENTER

ALARM

= ALARM1 (or ALARM2)

CALIBRATION = CAL
HOLD

= UP and DOWN ARROWS

DIAGNOSTICS = DISPLAY+ACK (from HOLD)