0 calibration and adjustments, 1 frequency of calibration, 2 calibration mode – Emerson 7003M User Manual

Page 47: A. zero calibration, B. span calibration

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Calibration and Adjustments 4-1

SECTION 4

CALIBRATION AND ADJUSTMENTS

4-1 FREQUENCY OF CALIBRATION

When a new sensor is installed, adjust the
zero offset current (Section 3-5d on page 3-
15) and calibrate the instrument with span
gas (4-2b on page 4-1. Span the
instrument weekly at first and note the
stability. Use this information to determine
the regular calibration frequency required
by this particular application.

4-2 CALIBRATION MODE

Refer to Figure 4-1 on page 4-3and Table
4-1 below. From the CALIBRATION Mode,
the operator can:

Calibrate with span gas

Enter sample and span gas pressures

KEY

DISPLAY

WHAT HAPPENS

DISPLAY

Nothing

Nothing

ACK

Nothing

Nothing

ALARM 1

Nothing

Nothing

ALARM 2

Nothing

Nothing

UP ARROW

10.8

→10.9

Changes calibration factor

DOWN ARROW

10.9

→10.8

Changes calibration factor

CAL

"PPr"

Moves to pressure compensation mode

ENTER

"run"

Drops back to RUN mode

Table 4-1.

CALIBRATION Mode Key Functions

a. Zero

Calibration

Each oxygen sensor produces a
constant residual electrical current in
the absence of oxygen due to
characteristics of the sensor. Typically
this current is less that 0.1% apparent
oxygen. It is different for each sensor,
but will remain constant for the life of
that sensor. The instrument can be
zeroed by adjusting the residual current
per Section 3-5d on page 3-15 After
this procedure is performed once, the
zero should not need adjusting again
during the life of the sensor.

b. Span

Calibration

1. Expose the sensor to ambient air

1

and let the oxygen reading and
sensor temperature stabilize.

1

We recommend using ambient air in calibration, re-

gardless of the range. Only when monitoring at 0 to

2. Press CAL to enter the

CALIBRATION mode. Enter the
security code if required.

3. CAL flashes and oxygen

concentration is displayed with the
right digit flashing. Output is held
at last reading until returned to
RUN mode.

4. Use the ARROW keys to set the

instrument to the concentration of
the calibration gas. See Table 4-2
on page 4-2 and Table 4-3 on page
4-2 if using ambient air as span
gas.

1% range, can the accuracy of the instrument be
slightly improved by using a gas of lower concentra-
tion (e.g., 1%) for calibration. If bottled air is used,
verify the oxygen concentration against National Bu-
reau of Standards certified calibration gas. The true
oxygen concentration in cylinders is often different
from the concentration indicated on the cylinder label.

Advertising