D. zero offset adjustment, Model 7003m – Emerson 7003M User Manual

Page 41

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-15

Model 7003M

d. Zero Offset Adjustment

Following the last procedure, the current
zero offset value will be flashing.

NOTE

When flowing zero gas over the sen-
sor, ensure that there are no leaks and
the gas actually has zero oxygen con-
centration. Otherwise, significant er-
rors will result. Zero offset adjustment
is generally not required. Errors are
more often due to incorrect zero gas
oxygen concentration than residual
current of the oxygen sensor.

The zero offset adjustment allows the
operator to compensate for any residual
sensor current. Adjustment range is 0-5%
equivalent oxygen.

Sensor Current = Current from
Oxygen + Residual Current

NOTE

Adjusting the zero offset affects the
output of the Model 7003M. Therefore,
if the zero offset is changed, the sys-
tem must be re-calibrated with span
gas.

Zero Offset Is Not To Be Changed

Press ENTER to return analyzer to RUN
mode, and skip the remainder of this
procedure.

Zero Offset Is To Be Removed

1. Press the UP ARROW until the

maximum value is displayed.

2. Press ENTER and re-calibrate with

span gas per Section 4-2b on page 4-
1.

Zero Offset To Be Adjusted

1. Expose the sensor to 100% oxygen-

free nitrogen or other inert gas and
allow the sensor signal to stabilize (2-
5 hours depending on sample
conditions).

2. Use the DOWN ARROW to bring the

displayed value to zero (for maximum
offset).

3. Press ENTER to return to RUN mode.

4. Re-calibrate with span gas per

Section 4-2b on page 4-1.

Advertising