4-4 Calibration and Adjustments

Rosemount Analytical Inc. A Division of Emerson Process Management

Model 7003M

BAROMETRIC

PRESSURE

CM HG (IN. HG)

APPARENT O

2

PARTIAL

PRESSURE

CM HG (DRY AIR)

O

2

CONCENTRATION WITH

NO PRESSURE CORRECTION

1

73 (28.7)

15.3

20.24%

74 (29.1)

15.5

20.52

75 (29.5)

15.7

20.80

76 (29.9)

15.8

20.95

77 (30.3)

16.1

21.36

78 (30.7)

16.3

21.63

1

Computed using 20.946% oxygen at 760 m.

Table 4-4.

Barometric Pressure Effect on Apparent Oxygen Concentration

KEY

DISPLAY

WHAT HAPPENS

DISPLAY

Nothing

Nothing

ACK

Nothing

Nothing

ALARM 1

Nothing

Nothing

ALARM 2

Nothing

Nothing

UP ARROW

10.8

→10.9

Changes pressure value

DOWN ARROW

10.9

→10.8

Changes pressure value

CAL

Nothing

Nothing

ENTER

"CPr" / "run"

Moves to calibration pressure input or goes back to RUN mode

Table 4-5.

PRESSURE COMPENSATION Mode Key Functions

c. Pressure Compensation - Correction

for Constant Pressure Differences

Refer to Figure 4-1 on page 4-3 and
Table 4-5 above. If the sample and
calibration gases are each at constant
pressure, the following procedure should
be used to correct for the pressure
difference between them. However, the
instrument does not monitor pressure and
will not correct the output if either sample
or calibration gas pressure changes.
(See Section 5-2 on page 5-1).

1. From RUN mode, press CAL.

2. CAL, YES, CAL flash briefly, then the

oxygen concentration is displayed
with the right digit flashing.

3. Press CAL again.

4. PPr flashes, then previous sample

pressure setting is displayed with right
digit flashing (76.0 cm Hg is default).

5. Use ARROW keys to enter sample

pressure.

6. Press ENTER to accept the value.

7. CPr flashes, then previous span gas

pressure setting is displayed with right
digit flashing (76.0 cm Hg is default).

8. Use the ARROW keys to enter

pressure of the calibration gas to be
used. If the sample and calibration
gas will always be at the same
pressure, set PPr and CPr at the
same value.

9. Press ENTER to accept the value.