I. setting alarm on and off levels, 6 action of alarm relay on and off setpoints, Model 7003m – Emerson 7003M User Manual

Page 45: Operator manual

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Operation 3-19

Model 7003M

ACTIVATES

DE-ACTIVATES

OFF SETPOINT

SETPOINT

ON

OFF

8.24%

8.10%

TIME

ALARM ON

DEADBAND

LOW Alarm

set for 8.10% O

2

ACTIVATES

DE-ACTIVATES

SETPOINT

OFF SETPOINT

ON

OFF

8.24%

8.10%

TIME

ALARM ON

HIGH Alarm

set for 8.24% O

2

i.

Setting Alarm ON and OFF Levels

Refer to flowchart Figure 3-5 on page 3-
17.

From RUN mode, press ALARM 1.

A-1, YES, A-1 (or C-1), ON flash briefly;
the alarm ON setpoint is displayed with
the right digit blinking.

A-1 indicates relay 1 acts as an ALARM;
C-1 indicates relay one acts as an
ON/OFF-Control.

To change the relay action from
ALARM to ON/OFF Control function
(or
vise versa) press ALARM 1. C-1 or A-1
will flash indicating a new mode.

To disable ALARM 1 - Press ACK.
"DIS" will flash briefly and return to RUN
mode.

To set the ON setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
OFF is flashed briefly, then the alarm OFF
setpoint is displayed with the right digit
flashing.

To set the OFF setpoint for ALARM 1 -
Use the ARROW keys, press ENTER.
The instrument displays "run" and returns
to RUN mode.

Figure 3-6. Action of Alarm Relay ON and OFF Setpoints