Operation 3-9

Model 7003M

b. System

Restart

The Model 7003M will attempt to restart
itself in the event of a power interruption,
power surge or other outside interference
to microprocessor operation (Section 3-3c
below).

The analyzer can also be manually
restarted if the RAM or EEPROM
becomes corrupted (Section 3-3d below).

c. Automatic Restart - Due to Power

Failure

If a power failure or other event occurs
that affects operation of the
microprocessor the system will attempt to
re-start itself. If it is successful, the
instrument will continue to operate
normally, except that the appropriate error
code will be displayed until the ACK
button in pressed. See Section 5-3c on
page 5-2 for circuit description.

d. Manual

Restart

1. Turn power to analyzer OFF.

2. Wait 30 seconds, and turn the power

to the analyzer ON.

3. While the software revision is being

displayed and the self-tests are being
performed, press DISPLAY and ACK
simultaneously.

"dYn" is displayed.

4. Press ENTER to clear the RAM and

EEPROM.

The display will go blank, then continue
with the SETUP routine.

Continue with Section 3-5a on page 3-13,
step 2.