A. single and double keystroke, 1 key functions (how to use), Model 7003m – Emerson 7003M User Manual

Page 29

Advertising
background image

Operator Manual

748223-K

June 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Replacement Parts 3-3

Model 7003M

a. Single and Double Keystroke

Most procedures involve a sequence of
single keystrokes by which the operator

presses one key at a time. Some of the
more frequently used single-key
commands are listed in Table 3-1 below.

KEY

FUNCTION

HOW TO USE IT

ACK

Acknowledges an alarm condition

Press once to silence alarms

DISPLAY

Switch between %O2 and

°C

Press to display

°C; press again to display O2.

Output signals are not affected.

ARROW keys

Increase/Decrease displayed value

Press once to change value by one digit. Hold key
down to change value quicker.

CAL

Access CALIBRATION mode

Press to calibrate. (If security is enabled, prompts
for ID code first)

ENTER

End command sequence

Press to enter value or to end a command se-
quence

ALARM1

Access ALARM 1 mode

Press to check, set or disable Alarm 1 or ON/OFF-
Control relays

ALARM2

Access ALARM 2 mode

Press to check, set or disable Alarm 2 or ON/OFF-
Control relays

CAL + ENTER

Access SETUP mode

From RUN mode, press CAL and ENTER keys
simultaneously.

UP + DOWN ARROW

Access HOLD mode.

From RUN mode, press UP ARROW and DOWN
ARROW keys simultaneously.

DISPLAY + ACK

Access DIAGNOSTICS mode

From HOLD mode, press DISPLAY and ACK keys
simultaneously.

DISPLAY + ACK

Access security ID code

From RUN mode, press DISPLAY and ACK keys
simultaneously.

Table 3-1.

Key Functions

INDICATOR

CARET/STATUS

DESCRIPTION

%O2

Caret ON

Oxygen percentage displayed

°C

Caret ON

Sensor temperature displayed,

°C

ALARM1,2

Caret OFF

Normal condition

ALARM1,2

Caret blinking

Alarm condition

-

Caret ON

Displays current output in mA
(DIAGNOSTICS mode only)

Table 3-2.

Display Indicators