5.7 Confirming Magnetron Heater Voltage

Magnetron heater voltage is adjusted at the factory. However, confirm that it is
within the prescribed rating as follows:

1. [MENU] [0] [0] [0] [0] [5] [5] [0] [0] [0] [0] to display the OTHER menu.

Figure 5-8 OTHER menu

2. Press the [3] key to select STOP from the TX/ANT field and press the [EN-

TER/SELECT] key.

3. Disconnect connector P821 from the scanner unit.

4. Measure on the 0.125 nm and 48 nm range scales, the voltage between pins

#12(+) and #5(-) on connector P801 on the RFC Board (03P9243) in the scan-
ner unit.

5. If the voltage is not within the rating shown in Table 5-2, adjust potentiometer

VR1 on the RFC Board.

Table 5-2 Magnetron ratings

Rating

FR-1505 MK3

(6 kw)

FR-1510 MK3

(12 kW)

FR-1525 MK3

(25 kW)

0.125 nm

7.4 - 7.6 V

7.4 - 7.6 V

8.2 - 8.4 V

48 nm

7.4 - 7.6 V

7.4 - 7.6 V

6.5 - 7.5 V

6. Set ROTATE from the TX/ANT field and press the [ENTER/SELECT] key.

Figure 5-9 RFC Board