Application parameters, Volume (p050 to p055), Application parameters volume (p050 to p055) – Siemens sitrans LU01 User Manual

Page 38: P050 tank shape

Advertising
background image

Page 32

SITRANS LU01 – INSTRUCTION MANUAL

7ML19985BE02

Pa

ra

m

e

te

rs

Application Parameters

Volume (P050 to P055)

Use these parameters to enable the SITRANS LU01 to show readings based on vessel
volume (rather than level).

P050 Tank Shape

Enter the Tank Shape value matching the monitored vessel or wet well.
When Operation is level (P001 = 1), liquid (material) volume is calculated. Alternatively,
when Operation is space (P001 = 2), remaining vessel capacity is calculated. In RUN
mode, readings are displayed in percent of maximum volume. To convert readings to
volumetric units, see

Maximum Volume (P051).

Values

#

Shape

Description

0

*

volume calculation not required
(preset)

1

Flat Level Bottom

2

Cone/Pyramid Bottom

3

Parabola Bottom

4

Half Sphere Bottom

5

Flat Sloped Bottom

6

Flat Ends