P902 watchdog, P903 display, P904 keypad – Siemens sitrans LU01 User Manual

Page 87: P905 transmit pulse, P906 communications

Advertising
background image

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 81

m

mmmm

Pa
ra

m

e

te

rs

P902 Watchdog

Press ENTER

to put the CPU into an infinite loop to test the watchdog timer.

On successful completion (10 seconds) the RUN mode is entered and the SITRANS LU01
is reset. Programming is kept and the unit responds as if there had been a power failure.

P903 Display

Press ENTER

to activate the display test.

All LCD segments and symbols are temporarily displayed.

P904 Keypad

Press ENTER

, then press each keypad key in the following sequence:

As each key is pressed, the associated keypad number is displayed. On successful test
completion, PASS is displayed. FAIL is displayed if a key is pressed out of sequence or the
programmer keypad malfunctions.

P905 Transmit Pulse

Used to monitor the transmit pulse with an oscilloscope connected to the transducer

terminals.

Press ENTER

to supply repeated transmit pulses, at the frequency entered, to the

transducer and / or view the transducer operating frequency (automatically altered by
P004 Transducer) for the Point Number displayed.

P906 Communications

Press ENTER

to test the SITRANS LU01 communications circuitry.

On successful completion, PASS is displayed. Otherwise FAIL is displayed. If FAIL is
displayed, repeate the test (the first test performed sets up the auto polarity function).

Values

Range: 10 to 60 kHz

Altered By

P004 Transducer

Note:

Before activating this feature, connect terminal block contacts 27 to 29

and 28 to 30.