SITRANS LU01 – INSTRUCTION MANUAL

Page 35

m

mmmm

Pa
ra

m

e

te

rs

Entering a Level Breakpoint

1.

Go to Parameter P054.

2.

Press

to display the index symbol.

3.

For each index enter a volume.

4.

Ensure that each volume corresponds to the same index for P055.

5.

Press .

For more on Volume Characterization, go to page 87.

P055 Volume Breakpoints and Characterization

(Universal Volume Calculation)

Each segment defined by the level breakpoints (P055) requires a volume so that the

SITRANS LU01 can make the level-to-volume calculations.

Typical volume calculations

Entering a Volume Breakpoint

1.

Go to Parameter P055.

2.

Press

to display the index symbol.

3.

For each index enter a volume.

4.

Ensure that each volume corresponds to the same index for P054.

5.

Press .

For more on Volume Characterization, go to page 87.

Primary Index

Breakpoint

Values

Range: 0.0 to 9999

Related

P054 Breakpoints Levels (Universal Volume Calculation)

Cone

Cylinder

V = (1/3)πr

2

h

V = πr

2

h