Siemens sitrans LU01 User Manual

Page 99

Advertising
background image

7ML19985BE02

SITRANS LU01 – INSTRUCTION MANUAL

Page 93

mm

mmm

Te

c

hnical Ref

e

re

nce

Parameter Settings for Point # 1

Parameter Settings for Ma Output #1

Parameter Setting for Temperature Averaging

With process control equipment disabled, enter the RUN mode to monitor system
performance and alarm programming. (A Reading Measurement simulation (P920) may
be performed to verify mA output programming).

When satisfied with system performance and programming, connect mA output 1 to the
Programmable Logic Controller (PLC).

Note that on power off the mA output drops to 0.00 mA. During a power interruption, the
last known mA output value is maintained in memory for 1 hour minimum. If power is
interrupted beyond 1 hour, on power resumption, the mA output immediately assumes the
new value.

Number

Description

Instructions

P001

Operation

Enter "2" for

space

.

P002

Material

No entry required (preset for

liquid)

.

P003

Measurement Response

Enter "4" for

surge

(100 m/min, Failsafe Timer = 0.1

minutes).

P004

Transducer

Enter "101" for

XCT-8

.

P005

Units

Enter "4" for

feet

.

P006

Empty

Enter "4.5" for

4.50 ft

(3 feet diameter plus 1.5 ft in

standpipe).

P007

Span

Enter "3" for

3 ft

(vessel diameter).

P070

Failsafe Timer

No entry required, (preset by P003).

Number

Description

Instructions

P200

mA Range

No entry required. (preset for

20 mA = low level

,

P001 = 2.)

P213

mA Max Limit

Enter "22" for

22.00 mA

.

P219

mA Failsafe

Enter "20.2" for

20.20 mA

.

Number

Description

Instructions

P660

Temp Source

Enter "5" for average XCT-8 and TS-3 temperatures.

Advertising