Chapter 3

Signal Connections

PCI-1200 User Manual

3-10

© National Instruments Corporation

Differential Connections for Grounded Signal Sources

Figure 3-3 shows how to connect a ground-referenced signal source to a
PCI-1200 board configured for DIFF input. Configuration instructions are
in the

Analog I/O Configuration

section in Chapter 2,

Installation and

Configuration

.

Figure 3-3. Differential Input Connections for Grounded Signal Sources

With this type of connection, the instrumentation amplifier rejects both the
common-mode noise in the signal and the ground-potential difference
between the signal source and the PCI-1200 ground (shown as V

cm

in

Figure 3-3).

+

+

Grounded

Signal

Source

V

m

Measured

Voltage

V

s

+

I/O Connector

PCI-1200 in DIFF Configuration

1

3

5

7

2

4

6

8

Common-

Mode

Noise,

Ground

Potential

ACH0

AGND

11

ACH2

ACH4

ACH6

ACH1

ACH3

ACH5

ACH7

V

cm

+

PC0